FAQs

How can I collect Pressure and Temperature data in the same software?

Setting the data collection window with a map has always been the first configuration step in I-Scan. If you've used it before, this will be very familiar. By selecting a Temperature map, the software knows to enable temperature features for that window. The software will also enable the Temperature legend which uses a unique color scale from the pressure map to ease the analysis.

 

Link to this FAQ

What is the data capture rate of the temperature mapping system?

The Evolution handle used in the I-Scan temperature mapping system can capture data on the sensor at up to 100 Hz.

 

Link to this FAQ

How do I clean my T-Scan system components?

Question: How do I clean my T-Scan™ system components?

Answer: The T-Scan III scanning handle cannot be sterilized but may be wrapped in plastic wrap for isolation from the patient. The sensor supports can be autoclaved or placed in cold sterilization liquid. Both the sensors and sensor supports can be cleaned with alcohol swabs. The same sensor can be used on the same patient across multiple sessions. The same sensor should not be used on different patients. Sensors are disposable after their useful life. Research has proven that our sensors can be used approximately 15-25 times each while still retaining our high standards of reliability.

Link to this FAQ