FAQs

Can Tekscan sensors be cut or trimmed?

Question: Can Tekscan sensors be cut or trimmed?

Answer: Yes, some of our sensors can be cut or trimmed to smaller sizes or shapes, or even punctured for secure placement. The model 4205 (pictured), model 7800 pressure sensor and T800 temperature sensor are designed to be sized for custom embedding in battery placements or other tight spots.

image-20240820133832-1Tips for trimming:

The rows and columns are connected through silver dots and traces on the surface of the sensor. If you trim the sensor, a row or column will remain active so long as the silver dot and silver connecting trace remain intact and connected to the handle.

Based on this photo, you can trim up to 15 rows from the distal end of the sensor before you reach the column trace connections. And you can trim up to 5 columns from either side before you reach the silver connections to the rows.

Consult with a Tekscan engineer if you aren't sure if your sensor is trimmable or how to do it. Cutting through the sensing element will short the sensor! 


---
The ability to trim for shoe size is one of the most important advantages of our in-shoe F-Scan™GO sensor, which can be trimmed to fit any shoe size up to a men’s size 14.

Tips for trimming the F-Scan GO sensor can be found in the manual, or HERE. 

Link to this FAQ

Error: Handle Driver is Not Loading - Error Code 52

Issue: T-Scan or I-Scan doesn't see the Evolution handle. The Device Manager shows Code 52 error.

T-Scan 9.1 Handle Driver Error

Solution:

First enable Windows Updates on your system. Install all the critical updates and see if this fixes the Code 52 error.

If not, this could be caused by a required security patch from Microsoft (that is missing). Info on the update needed:

https://support.microsoft.com/en-us/kb/3033929

Download of the update:

https://technet.microsoft.com/en-us/library/security/3033929

"SHA-2 certificates require KB3033929 on Windows 7

If your certificate uses SHA-2 or has SHA-2 certificates in its chain of trust and you are using it to sign kernel modules, then you should be aware of KB3033929, an update for Windows 7 distributed through Windows Update. On versions of Windows 7 without this update, the kernel will reject signatures made with certificates that use SHA-2, so they cannot be used to get a kernel module to load."

Select your Windows OS, download, install and re-boot the computer.

Link to this FAQ

What is calibration?

Question: What is calibration?

Answer: Calibration is the method by which the digital output of the system is converted to an actual engineering unit. Calibration enables you to compare the output of the same sensor in various environments and allows you to compare calibrated outputs of various sensors.

Link to this FAQ