Can I do a one-point calibration with ambient temperature?

No, the slope of the temperature sensor’s calibration curve does not run through zero.  Two points are required to calibrate the sensor.  If a single point calibration is performed the calibration point will be saved in the system, but no calibration will be applied to the sensor.

 