The Real-time Status Bar reads ‘MISALIGNED!’ instead of ‘Sensor OK’

Issue: The Real-time Status Bar reads ‘MISALIGNED!’ instead of ‘Sensor OK’.

Solution: possible causes and actions to take:

 

  • Sensor is misaligned or was pulled out of the handle.

    Remove and re-insert Sensor.

     

  • Sensor is inserted upside down in the handle.

    Remove the Sensor.  Make certain that the This Side UP legend is facing the handle latch.

     

  • The terminals on the Sensor are dirty.

    Clean the terminal sections of the Sensor carefully with a cloth moistened with alcohol.

     

  • Two Sensors are stuck together.

    Remove Sensors and peel apart.  Re-insert one Sensor.

     

  • Sensor is damaged.

    Replace the Sensor with a different Sensor.

     

  • The Handle electronics are damaged.

    If multiple Sensors produce the same “MISALIGNED” error message, there may be a problem with the electronics.  Contact Tekscan Technical Support.

     

Revision Date: 08/15/2007

Did this Article Solve your Issue?

If you did not find a complete solution to your problem, please let us know by filling out the following form or calling us directly.

Tekscan Technical Support Hours
Monday - Friday: 8:00am - 7:00pm EST, Excluding Holidays

To Reach Tekscan Technical Support

Call (800) 248-3669 x359 Toll Free within the US and Canada.
Call 1(617) 464-4500 x359 International
Call 1(617) 464-4266 International
Email: support@tekscan.com

We respond promptly to customer problems, and new support articles are created from your input.

Name: Email:
Hardware Serial #: ** Software Serial #:
Describe your Issue:


Mollom CAPTCHA


Please enter the value of the security question above.



* - The software serial number can be found in the Help drop down menu of the program. Navigate to Help -> About.

** - The hardware serial number can be found on the back of each Handle/Cuff/Hub preceded by S/N.