The Real-time Status Bar reads ‘MISALIGNED!’ instead of ‘Sensor OK’
Issue: The Real-time Status Bar reads ‘MISALIGNED!’ instead of ‘Sensor OK’.
Solution: possible causes and actions to take:
- Sensor is misaligned or was pulled out of the handle.
Remove and re-insert Sensor. - Sensor is inserted upside down in the handle.
Remove the Sensor. Make certain that the This Side UP legend is facing the handle latch. - The terminals on the Sensor are dirty.
Clean the terminal sections of the Sensor carefully with a cloth moistened with alcohol. - Two Sensors are stuck together.
Remove Sensors and peel apart. Re-insert one Sensor. - Sensor is damaged.
Replace the Sensor with a different Sensor. - The Handle electronics are damaged.
If multiple Sensors produce the same “MISALIGNED” error message, there may be a problem with the electronics. Contact Tekscan Technical Support.